BELLEZZO, MURILLO
; FONSECA, GABRIEL P.; VONCKEN, ROBERT; VERRIJSSEN, AN-SOFIE; VAN BEVEREN, CELINE; ROELOFS, ERIK; YORIYAZ, HELIO
; RENIERS, BRIGITTE; VAN LIMBERGEN, EVERT J.; BERB EE, MAAIKE; VERHAEGEN, FRANK.
Advanced design, simulation, and dosimetry of a novel rectal applicator for contact brachytherapy with a conventional HDR 192Ir source.
Brachytherapy,
v. 19,
n. 4,
p. 544-553,
2020.
DOI:
10.1016/j.brachy.2020.03.009
Abstract:
PURPOSE: Dose escalation yields higher complete response to rectal tumors, which may enable
the omission of surgery. Dose escalation using 50 kVp contact x-ray brachytherapy (CXB) allow the
treatment of a selective volume, resulting in low toxicity and organs-at-risk preservation. However,
the use of CXB devices is limited because of its high cost and lack of treatment planning tools.
Hence, the MAASTRO applicator (for HDR 192Ir sources) was developed and characterized by
measurements and Monte Carlo simulations to be a cost-effective alternative to CXB devices.
METHODS AND MATERIALS: A cylindrical applicator with lateral shielding was designed to
be used with a rectoscope using its tip as treatment surface. Both the applicator and the rectoscope
have a slanted edge to potentially allow easier placement against tumors. The applicator design was
achieved by Monte Carlo modeling and validated experimentally with film dosimetry, using the
Papillon 50 (P50) device as reference.
RESULTS: The applicator delivers CXB doses in less than 9 min using a 20375 U source for a
treatment area of approximately 20 20 mm2 at 2 mm depth. Normalized at 2 mm, the dose falloff
for depths of 0 mm, 5 mm, and 10 mm are 130%, 70%, and 43% for the P50 and 140%, 67%, and
38% for the MAASTRO applicator, respectively.
CONCLUSIONS: The MAASTRO applicator was designed to use HDR 192Ir sources to deliver a
dose distribution similar to those of CXB devices. The applicator may provide a cost-effective solution
for endoluminal boosting with clinical treatment planning system integration.
Palavras-Chave:
brachytherapy;
monte carlo method;
rectal administration;
iridium 192;
dose rates;
radiation sources;
simulation;
radiotherapy;
neoplasms;
film dosimetry
BELLEZZO, MURILLO; FONSECA, GABRIEL P.; VONCKEN, ROBERT; VERRIJSSEN, AN-SOFIE; VAN BEVEREN, CELINE; ROELOFS, ERIK; YORIYAZ, HELIO; RENIERS, BRIGITTE; VAN LIMBERGEN, EVERT J.; BERB EE, MAAIKE; VERHAEGEN, FRANK.
Advanced design, simulation, and dosimetry of a novel rectal applicator for contact brachytherapy with a conventional HDR 192Ir source.
Brachytherapy,
v. 19,
n. 4,
p. 544-553,
2020.
DOI:
10.1016/j.brachy.2020.03.009.
Disponível em: http://repositorio.ipen.br/handle/123456789/31351. Acesso em: $DATA.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.
Como referenciar este item
-
FONSECA, GABRIEL P.
; LUVIZOTTO, JESSICA
; ANTUNES, PAULA C.G.
; YORIYAZ, HELIO
.
Dose measurements for an Ir-192 source using mosfets, radiochromic films, TLDs and heterogeneous tissues.
Brachytherapy,
v. 13,
p. S103,
2014.
DOI:
10.1016/j.brachy.2014.02.388
Palavras-Chave:
brachytherapy;
iridium 192;
thermoluminescent dosimetry;
accuracy;
animal tissues
FONSECA, GABRIEL P.; LUVIZOTTO, JESSICA; ANTUNES, PAULA C.G.; YORIYAZ, HELIO.
Dose measurements for an Ir-192 source using mosfets, radiochromic films, TLDs and heterogeneous tissues.
Brachytherapy,
v. 13,
p. S103,
2014.
DOI:
10.1016/j.brachy.2014.02.388.
Disponível em: http://repositorio.ipen.br/handle/123456789/23866. Acesso em: $DATA.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.
Como referenciar este item
-
BELLEZZO, MURILLO
; BAEZA, JOSE A.; VONCKEN, ROBERT; RENIERS, BRIGITTE; VERHAEGEN, FRANK; FONSECA, GABRIEL P..
Mechanical evaluation of the Bravos afterloader system for HDR brachytherapy.
Brachytherapy,
v. 18,
n. 6,
p. 852-862,
2019.
DOI:
10.1016/j.brachy.2019.06.005
Abstract:
PURPOSE: The Bravos afterloader system was released by Varian Medical Systems in October of
2018 for high-dose-rate brachytherapy with 192Ir sources, containing new features such as the
CamScale (a new device for daily quality assurance and system recalibration), channel length verification,
and different settings for rigid and flexible applicators. This study mechanically evaluated
the Bravos system precision and accuracy for clinically relevant scenarios, using dummy sources.
METHODS AND MATERIALS: The system was evaluated after three sets of experiments: (1)
The CamScale was used to verify inter- and intra-channel dwelling variability and system calibration;
(2) A high-speed camera was used to verify the source simulation cable movement inside a
transparent quality assurance device, where dwell positions, dwell times, transit times, speed profiles,
and accelerations were measured; (3) The source movement inside clinical applicators was
captured with an imaging panel while being exposed to an external kV source. Measured and
planned dwell positions and times were compared.
RESULTS: Maximum deviations between planned and measured dwell positions and times for the
source cable were 0.4 mm for the CamScale measurements and 0.07 seconds for the high-speed
camera measurements. Mean dwell position deviations inside clinical applicators were below
1.2 mm for all applicators except the ring that required an offset correction of 1 mm to achieve
a mean deviation of 0.4 mm.
CONCLUSIONS: Features of the Bravos afterloader system provide a robust and precise treatment
delivery. All measurements were within manufacturer specifications.
Palavras-Chave:
afterloading;
irradiation procedures;
brachytherapy;
calibration;
dose rates;
dosimetry;
radiation dose distributions
BELLEZZO, MURILLO; BAEZA, JOSE A.; VONCKEN, ROBERT; RENIERS, BRIGITTE; VERHAEGEN, FRANK; FONSECA, GABRIEL P.
Mechanical evaluation of the Bravos afterloader system for HDR brachytherapy.
Brachytherapy,
v. 18,
n. 6,
p. 852-862,
2019.
DOI:
10.1016/j.brachy.2019.06.005.
Disponível em: http://repositorio.ipen.br/handle/123456789/30451. Acesso em: $DATA.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.
Como referenciar este item
-
FONSECA, GABRIEL P.
; RENIERS, BRIGITTE; LANDRY, GUILLAUME; WHITE, SHANE; BELLEZZO, MURILLO
; ANTUNES, PAULA C.G.
; SALES, CAMILA P. de; WELTEMAN, EDUARDO; YORIYAZ, HELIO
; VERHAEGEN, FRANK.
A medical image-based graphical platform-Features, applications and relevance for brachytherapy.
Brachytherapy,
v. 13,
n. 6,
p. 632-639,
2014.
Palavras-Chave:
brachytherapy;
images;
image processing;
monte carlo method;
radiation doses;
calculation methods;
algorithms;
simulation;
computer codes
FONSECA, GABRIEL P.; RENIERS, BRIGITTE; LANDRY, GUILLAUME; WHITE, SHANE; BELLEZZO, MURILLO; ANTUNES, PAULA C.G.; SALES, CAMILA P. de; WELTEMAN, EDUARDO; YORIYAZ, HELIO; VERHAEGEN, FRANK.
A medical image-based graphical platform-Features, applications and relevance for brachytherapy.
Brachytherapy,
v. 13,
n. 6,
p. 632-639,
2014.
Disponível em: http://repositorio.ipen.br/handle/123456789/23185. Acesso em: $DATA.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.
Como referenciar este item
-
BELLEZZO, MURILLO
; FONSECA, GABRIEL P.; VERRIJSSEN, AN-SOFIE; VONCKEN, ROBERT; BOSCH, MICHIEL R.V. den; YORIYAZ, HELIO
; RENIERS, BRIGITTE; BERBEE, MAAIKE; LIMBERGEN, EVERT J.V.; VERHAEGEN, FRANK.
A novel rectal applicator for contact radiotherapy with HDR 192Ir sources.
Brachytherapy,
v. 17,
n. 6,
p. 1037-1044,
2018.
DOI:
10.1016/j.brachy.2018.07.012
Abstract:
PURPOSE: Dose escalation to rectal tumors leads to higher complete response rates and may
thereby enable omission of surgery. Important advantages of endoluminal boosting techniques
include the possibility to apply a more selective/localized boost than using external beam radiotherapy.
A novel brachytherapy (BT) rectal applicator with lateral shielding was designed to be used
with a rectoscope for eye-guided positioning to deliver a dose distribution similar to the one of contact
x-ray radiotherapy devices, using commonly available high-dose-rate 192Ir BT sources.
METHODS AND MATERIALS: A cylindrical multichannel BT applicator with lateral shielding
was designed by Monte Carlo modeling, validated experimentally with film dosimetry and
compared with results found in the literature for the Papillon 50 (P50) contact x-ray radiotherapy
device regarding rectoscope dimensions, radiation beam shape, dose fall-off, and treatment time.
RESULTS: The multichannel applicator designed is able to deliver 30 Gy under 13 min with a
20350 U (5 Ci) source. The use of multiple channels and lateral shielding provide a uniform circular
treatment surface with 22 mm in diameter. The resulting dose fall-off is slightly steeper (maximum
difference of 5%) than the one generated by the P50 device with the 22 mm applicator.
CONCLUSIONS: A novel multichannel rectal applicator for contact radiotherapy with high-doserate
192Ir sources that can be integrated with commercially available treatment planning systems
was designed to produce a dose distribution similar to the one obtained by the P50 device.
Palavras-Chave:
radiotherapy;
large intestine;
rectum;
neoplasms;
monte carlo method;
brachytherapy;
dose rates;
iridium 192
BELLEZZO, MURILLO; FONSECA, GABRIEL P.; VERRIJSSEN, AN-SOFIE; VONCKEN, ROBERT; BOSCH, MICHIEL R.V. den; YORIYAZ, HELIO; RENIERS, BRIGITTE; BERBEE, MAAIKE; LIMBERGEN, EVERT J.V.; VERHAEGEN, FRANK.
A novel rectal applicator for contact radiotherapy with HDR 192Ir sources.
Brachytherapy,
v. 17,
n. 6,
p. 1037-1044,
2018.
DOI:
10.1016/j.brachy.2018.07.012.
Disponível em: http://repositorio.ipen.br/handle/123456789/29271. Acesso em: $DATA.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.
Como referenciar este item
-
VERRIJSSEN, AN-SOFIE; OPBROEK, THIRZA; BELLEZZO, MURILLO
; FONSECA, GABRIEL P.; VERHAEGEN, FRANK; GERARD, JEAN-PIERRE; MYINT, ARTHUR S.; LIMBERGEN, EVERT J.V.; BERBEE, MAAIKE.
A systematic review comparing radiation toxicity after various endorectal techniques.
Brachytherapy,
v. 18,
n. 1,
p. 71-86,
2019.
DOI:
10.1016/j.brachy.2018.10.001
Observação: Corrigendum anexado. Brachytherapy, v. 18, p. 427, 2019. DOI: 10.1016/j.brachy.2019.03.010
Abstract:
PURPOSE: A clinical complete response is seen after neoadjuvant chemoradiation for rectal tumors
in 15%e20% of patients. These patients can potentially be spared mutilating total mesorectal
excision surgery through a watch-and-wait policy. Recent studies show that dose escalation by a
radiation boost increases the clinical complete response rate. The boost dose to the tumor can be
administered through external beam radiotherapy or through internal radiotherapy using techniques
like contact therapy, low-dose-rate or high-dose-rate brachytherapy (BT). However, limited information
is available concerning treatment-related toxicity of these techniques. With this systematic
review, we aim to summarize and compare published data concerning acute and late toxicity after
contact X-ray therapy (CXT) and BT for rectal cancer.
METHODS AND MATERIALS/RESULTS: Thirty-eight studies reporting toxicity after endorectal
radiation techniques for rectal cancer were included, resulting in 3682 patients for analysis.
Direct comparison of toxicity by the different radiation modes was hampered by various combinations
of endorectal techniques, a lack of clear reporting of toxicity scores, dose prescription, technique
used, and treated volumes. $ Grade 3 rectal toxicity was reported in 2.9% of patients having
received only CXT; 6.3% of patients who received only BT had Grade 3 rectal toxicity, and BT also
caused Grade 3 urinary toxicity in 1 patient.
CONCLUSION: All techniques reported some$Grade 3 toxicity. Toxicity after CXTwas confined
to the rectum, whereas after BT, urogenital toxicity and skin toxicity were seen as well. Unfortunately,
few specific conclusions could be drawn regarding the dose-related risk of toxicity for the various
techniques due to nonuniform reporting strategies and missing information. To enable future comparisons
and improvements, the endorectal radiation field urgently needs consensus guidelines on dose
reporting, dose prescription, treatment volume specification, and toxicity reporting. 2018 American
Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
Palavras-Chave:
chemotherapy;
brachytherapy;
neoplasms;
rectum;
radiation doses;
toxicity;
intestines
VERRIJSSEN, AN-SOFIE; OPBROEK, THIRZA; BELLEZZO, MURILLO; FONSECA, GABRIEL P.; VERHAEGEN, FRANK; GERARD, JEAN-PIERRE; MYINT, ARTHUR S.; LIMBERGEN, EVERT J.V.; BERBEE, MAAIKE.
A systematic review comparing radiation toxicity after various endorectal techniques.
Brachytherapy,
v. 18,
n. 1,
p. 71-86,
2019.
DOI:
10.1016/j.brachy.2018.10.001.
Disponível em: http://repositorio.ipen.br/handle/123456789/30067. Acesso em: $DATA.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.
Como referenciar este item
-
ANTUNES, PAULA C.G.
; SIQUEIRA, PAULO de T.D.
; SHORTO, JULIAN M.B.
; YORIYAZ, HELIO
.
A versatile physical phantom design and construction for I-125 dose measurements and dose-to-medium determination.
Brachytherapy,
v. 22,
n. 1,
p. 80-92,
2023.
DOI:
10.1016/j.brachy.2022.10.005
Abstract:
PURPOSE: In this paper we present a phantom designed to provide conditions to generate set of “true” independent reference data as requested by TG-186, and mitigating the scarcity of experimental studies on brachytherapy validation. It was used to perform accurate experimental measurements of dose of 125I brachytherapy seeds using LiF dosimeters, with the objective of experimentally validating Monte Carlo (MC) calculations with model-based dose calculation algorithm (MBDCA). In addition, this work intends to evaluate a methodology to convert the experimental values from LiF into dose in the medium. METHODS AND MATERIALS: The proposed PMMA physical phantom features cavities to insert a LiF dosimeter and a 125I seed, adjusted in different configurations with variable thickness. Monte Carlo calculations performed with MCNP6.2 code were used to score the absorbed dose in the LiF and the dose conversion parameters. A sensitivity analysis was done to verify the source of possible uncertainties and quantify their impact on the results. RESULTS: The proposed phantom and experimental procedure developed in this work provided precise dose data within 5.68% uncertainty (k = 1). The achieved precision made it possible to convert the LiF responses into absorbed dose to medium and to validate the dose conversion factor methodology. CONCLUSIONS: The proposed phantom is simple both in design and as in its composition, thus achieving the demanded precision in dose evaluations due to its easy reproducibility of experimental setup. The results derived from the phantom measurements support the dose conversion methodology. The phantom and the experimental procedure developed here can be applied for other materials and radiation sources.
Palavras-Chave:
biological models;
phantoms;
brachytherapy;
iodine 125;
seeds;
lithium fluorides;
monte carlo method
ANTUNES, PAULA C.G.; SIQUEIRA, PAULO de T.D.; SHORTO, JULIAN M.B.; YORIYAZ, HELIO.
A versatile physical phantom design and construction for I-125 dose measurements and dose-to-medium determination.
Brachytherapy,
v. 22,
n. 1,
p. 80-92,
2023.
DOI:
10.1016/j.brachy.2022.10.005.
Disponível em: http://repositorio.ipen.br/handle/123456789/34007. Acesso em: $DATA.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.
Como referenciar este item
Buscar no repositório
Navegar
Minha conta
Visualizar
A pesquisa no RD utiliza os recursos de busca da maioria das bases de dados. No entanto algumas dicas podem auxiliar para obter um resultado mais pertinente.
✔ É possível efetuar a busca de um autor ou um termo em todo o RD, por meio do
Buscar no Repositório
, isto é, o termo solicitado será localizado em qualquer campo do RD. No entanto esse tipo de pesquisa não é recomendada a não ser que se deseje um resultado amplo e generalizado.
✔ A pesquisa apresentará melhor resultado selecionando um dos filtros disponíveis em
Navegar
✔ Os filtros disponíveis em
Navegar
tais como: Coleções, Ano de publicação, Títulos, Assuntos, Autores, Revista, Tipo de publicação são autoexplicativos. O filtro,
Autores IPEN
apresenta uma relação com os autores vinculados ao IPEN; o
ID Autor IPEN
diz respeito ao número único de identificação de cada autor constante no RD e sob o qual estão agrupados todos os seus trabalhos independente das variáveis do seu nome;
Tipo de acesso
diz respeito à acessibilidade do documento, isto é , sujeito as leis de direitos autorais, ID RT apresenta a relação dos relatórios técnicos, restritos para consulta das comunidades indicadas.
A opção
Busca avançada
utiliza os conectores da lógica boleana, é o melhor recurso para combinar chaves de busca e obter documentos relevantes à sua pesquisa, utilize os filtros apresentados na caixa de seleção para refinar o resultado de busca. Pode-se adicionar vários filtros a uma mesma busca.
Exemplo:
Buscar os artigos apresentados em um evento internacional de 2015, sobre loss of coolant, do autor Maprelian.
Autor: Maprelian
Título: loss of coolant
Tipo de publicação: Texto completo de evento
Ano de publicação: 2015
✔ Para indexação dos documentos é utilizado o Thesaurus do INIS, especializado na área nuclear e utilizado em todos os países membros da
International Atomic Energy Agency – IAEA
, por esse motivo, utilize os termos de busca de assunto em inglês; isto não exclui a busca livre por palavras, apenas o resultado pode não ser tão relevante ou pertinente.
✔ 95% do RD apresenta o texto completo do documento com livre acesso, para aqueles que apresentam o
significa que e o documento está sujeito as leis de direitos autorais, solicita-se nesses casos contatar a Biblioteca do IPEN,
bibl@ipen.br
.
✔ Ao efetuar a busca por um autor o RD apresentará uma relação de todos os trabalhos depositados no RD. No lado direito da tela são apresentados os coautores com o número de trabalhos produzidos em conjunto bem como os assuntos abordados e os respectivos anos de publicação agrupados.
✔ O RD disponibiliza um quadro estatístico de produtividade, onde é possível visualizar o número dos trabalhos agrupados por tipo de coleção, a medida que estão sendo depositados no RD.
✔ Na página inicial nas referências são sinalizados todos os autores IPEN, ao clicar nesse símbolo
será aberta uma nova página correspondente à aquele autor – trata-se da página do pesquisador.
✔ Na página do pesquisador, é possível verificar, as variações do nome, a relação de todos os trabalhos com texto completo bem como um quadro resumo numérico; há links para o Currículo Lattes e o Google Acadêmico ( quando esse for informado).
ATENÇÃO!
ESTE TEXTO "AJUDA" ESTÁ SUJEITO A ATUALIZAÇÕES CONSTANTES, A MEDIDA QUE NOVAS FUNCIONALIDADES E RECURSOS DE BUSCA FOREM SENDO DESENVOLVIDOS PELAS EQUIPES DA BIBLIOTECA E DA INFORMÁTICA.
O gerenciamento do Repositório está a cargo da Biblioteca do IPEN. Constam neste RI, até o presente momento 20.950 itens que tanto podem ser artigos de periódicos ou de eventos nacionais e internacionais, dissertações e teses, livros, capítulo de livros e relatórios técnicos. Para participar do RI-IPEN é necessário que pelo menos um dos autores tenha vínculo acadêmico ou funcional com o Instituto. Nesta primeira etapa de funcionamento do RI, a coleta das publicações é realizada periodicamente pela equipe da Biblioteca do IPEN, extraindo os dados das bases internacionais tais como a Web of Science, Scopus, INIS, SciElo além de verificar o Currículo Lattes. O RI-IPEN apresenta também um aspecto inovador no seu funcionamento. Por meio de metadados específicos ele está vinculado ao sistema de gerenciamento das atividades do Plano Diretor anual do IPEN (SIGEPI). Com o objetivo de fornecer dados numéricos para a elaboração dos indicadores da Produção Cientifica Institucional, disponibiliza uma tabela estatística registrando em tempo real a inserção de novos itens. Foi criado um metadado que contém um número único para cada integrante da comunidade científica do IPEN. Esse metadado se transformou em um filtro que ao ser acionado apresenta todos os trabalhos de um determinado autor independente das variáveis na forma de citação do seu nome.
A elaboração do projeto do RI do IPEN foi iniciado em novembro de 2013, colocado em operação interna em julho de 2014 e disponibilizado na Internet em junho de 2015. Utiliza o software livre Dspace, desenvolvido pelo Massachusetts Institute of Technology (MIT). Para descrição dos metadados adota o padrão Dublin Core. É compatível com o Protocolo de Arquivos Abertos (OAI) permitindo interoperabilidade com repositórios de âmbito nacional e internacional.
1. Portaria IPEN-CNEN/SP nº 387, que estabeleceu os princípios que nortearam a criação do RDI,
clique aqui.
2. A experiência do Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN/SP) na criação de um Repositório Digital Institucional – RDI,
clique aqui.
O Repositório Digital do IPEN é um equipamento institucional de acesso aberto, criado com o objetivo de reunir, preservar, disponibilizar e conferir maior visibilidade à Produção Científica publicada pelo Instituto, desde sua criação em 1956.
Operando, inicialmente como uma base de dados referencial o Repositório foi disponibilizado na atual plataforma, em junho de 2015. No Repositório está disponível o acesso ao conteúdo digital de artigos de periódicos, eventos, nacionais e internacionais, livros, capítulos, dissertações, teses e relatórios técnicos.
A elaboração do projeto do RI do IPEN foi iniciado em novembro de 2013, colocado em operação interna em julho de 2014 e disponibilizado na Internet em junho de 2015. Utiliza o software livre Dspace, desenvolvido pelo Massachusetts Institute of Technology (MIT). Para descrição dos metadados adota o padrão Dublin Core. É compatível com o Protocolo de Arquivos Abertos (OAI) permitindo interoperabilidade com repositórios de âmbito nacional e internacional.
O gerenciamento do Repositório está a cargo da Biblioteca do IPEN. Constam neste RI, até o presente momento 20.950 itens que tanto podem ser artigos de periódicos ou de eventos nacionais e internacionais, dissertações e teses, livros, capítulo de livros e relatórios técnicos. Para participar do RI-IPEN é necessário que pelo menos um dos autores tenha vínculo acadêmico ou funcional com o Instituto. Nesta primeira etapa de funcionamento do RI, a coleta das publicações é realizada periodicamente pela equipe da Biblioteca do IPEN, extraindo os dados das bases internacionais tais como a Web of Science, Scopus, INIS, SciElo além de verificar o Currículo Lattes. O RI-IPEN apresenta também um aspecto inovador no seu funcionamento. Por meio de metadados específicos ele está vinculado ao sistema de gerenciamento das atividades do Plano Diretor anual do IPEN (SIGEPI). Com o objetivo de fornecer dados numéricos para a elaboração dos indicadores da Produção Cientifica Institucional, disponibiliza uma tabela estatística registrando em tempo real a inserção de novos itens. Foi criado um metadado que contém um número único para cada integrante da comunidade científica do IPEN. Esse metadado se transformou em um filtro que ao ser acionado apresenta todos os trabalhos de um determinado autor independente das variáveis na forma de citação do seu nome.